Model Number 2451-0007
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/31/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No product will be returned per customer.The customer complaint could not be confirmed because the product was not returned for failure investigation.The root cause of this failure was not identified.
 
Event or Problem Description
The customer reported that a mainline of lactated ringer was programmed at 20/hr.During an endoscopy procedure.The patient received iv sedation of midazolam and fentanyl and then was given a manual flush 10ml normal saline.Approximately 30 min later the user noticed set "blew apart¿ at an unspecified location.There was no report of patient harm.
 
Event or Problem Description
The customer reported that a mainline of lactated ringers was programmed at 20/hr (presumed to mean 20 ml/hr) during an endoscopy procedure.The patient received iv sedation of midazolam and fentanyl and then was given a manual flush of 10ml normal saline.Approximately 30 minutes later the user noticed that the set "blew apart¿ at an unspecified location.There was no report of patient harm.
 
Additional Manufacturer Narrative
The customer¿s report that the set came apart was confirmed.Visual inspection showed a separation between the silicone pump segment tubing and the upper fitment.The ring retainer was not received with the set.A ring retainer indentation was observed around the end of the silicone pump tubing where the separation occurred.The silicone tubing was measured and found to be within specification.Functional testing could not be performed due to the separation.The root cause of the separation could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS PUMP MODULE ADMINISTRATION SET
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
MDR Report Key7910729
Report Number9616066-2018-01750
Device Sequence Number1449813
Product Code FPA
UDI-Device Identifier10885403220432
UDI-Public10885403220432
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K944320
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 09/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/11/2021
Device Model Number2451-0007
Device Catalogue Number2451-0007
Device Lot Number18015780
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/12/2018
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date09/26/2018
Supplement Report FDA Received Date12/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
10ML SYRINGE, THERAPY DATE (B)(6) 2018
-
-