Model Number 7035M-90
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code Arthralgia (2355)
Date of Event 08/28/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Based on the information provided, review of the surgical technique manual, ifu, certificates of analysis and fmeas, there is no indication of device failure and no indication that the device was out of specification.The most probable root cause is user error; improper radiographic evaluation.The patient's dysmorphic sacrum may have contributed to the implant malpositioning.
 
Event or Problem Description
The patient had left side si joint arthrodesis in (b)(6) 2018 where three implants were placed.The patient has a dysmorphic sacrum.During the procedure, the caudal positioned implant was inadvertently advanced into the pelvis by the installation of the middle implant.The patient did not report any pain complaints.Later that same day, a revision procedure was performed where the malpositioned caudal implant was removed for safety reasons only as there were no pain complaints.No other preexisting implants were adjusted or removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IFUSE IMPLANT SYSTEM
Common Device Name
ORTHOPEDIC ROD
Manufacturer (Section D)
SI-BONE, INC.
471 el camino real
suite 101
santa clara CA 95050 4482
MDR Report Key7911482
Report Number3007700286-2018-00077
Device Sequence Number14029701
Product Code OUR
UDI-Device Identifier00851085007305
UDI-Public00851085007305
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K080398
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/02/2023
Device Model Number7035M-90
Device Lot Number2631321
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/28/2018
Initial Report FDA Received Date09/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age50 YR
-
-