Model Number 7050-90
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code Undesired Nerve Stimulation (1980)
Date of Event 04/24/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This previously unreported revision was discovered in an unrelated analysis of records.Based on the information provided, review of the surgical technique manual, ifu, certificates of analysis and fmeas, there is no indication of device failure and no indication that the device was out of specification.The most probable root cause is user error; improper implant size selection, using an implant that was too long.
 
Event or Problem Description
This previously unreported revision was discovered in an unrelated analysis of records.The patient had right side si joint arthrodesis in (b)(6) 2016 where three implants were placed.It is not known if the patient had any pain complaints.In (b)(6) 2018, the surgeon thought that the cephalad positioned implant may have been too close to the neuroforamen and decided to remove it.In (b)(6) 2018, the surgeon performed a revision procedure where he first removed the cephalad implant, likely using chisels.He then attempted to place an iliosacral screw in the explant void but the screw broke and was subsequently removed leaving the screw washer unnoticed in the void.After placing a shorter implant of the same initial type into the void, the surgeon noticed the washer on interoperative x-rays, removed the implant and the washer and replaced the implant.He finally added an additional implant of the same type in a ventral position.No other preexisting implants were adjusted or removed.The status of the patient following the revision procedure is not known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IFUSE IMPLANT SYSTEM
Common Device Name
ORTHOPEDIC ROD
Manufacturer (Section D)
SI-BONE, INC.
471 el camino real
suite 101
santa clara CA 95050 4482
MDR Report Key7911570
Report Number3007700286-2018-00079
Device Sequence Number13802831
Product Code OUR
UDI-Device Identifier00859256003309
UDI-Public00859256003309
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K080398
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/22/2020
Device Model Number7050-90
Device Lot Number493594
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/24/2018
Initial Report FDA Received Date09/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age52 YR
-
-