SI-BONE, INC. IFUSE IMPLANT SYSTEM; ORTHOPEDIC ROD Back to Search Results
Model Number 7060-90
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code Undesired Nerve Stimulation (1980)
Date of Event 06/04/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This previously unreported revision was discovered in an unrelated analysis of records.Based on the information provided, review of the surgical technique manual, ifu, certificates of analysis and fmeas, there is no indication of device failure and no indication that the device was out of specification.The most probable root cause is user error; improper implant size selection, using an implant that was too long.
 
Event or Problem Description
This previously unreported revision was discovered in an unrelated analysis of records.The patient had right side si joint arthrodesis in (b)(6) 2017 where two implants were placed.It is not known if the patient had any pain complaints.In (b)(6) 2018, the surgeon thought that the superior positioned implant may have been encroaching on the neuroforamen and decided to remove it.In (b)(6) 2018, the surgeon performed a revision procedure where he first removed the superior positioned implant using chisels.He then added an additional implant of the same type in a more inferior position.No other preexisting implants were adjusted or removed.The status of the patient following the revision procedure is not known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IFUSE IMPLANT SYSTEM
Common Device Name
ORTHOPEDIC ROD
Manufacturer (Section D)
SI-BONE, INC.
471 el camino real
suite 101
santa clara CA 95050 4482
MDR Report Key7911675
Report Number3007700286-2018-00084
Device Sequence Number16659988
Product Code OUR
UDI-Device Identifier00859256003323
UDI-Public00859256003323
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K080398
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/01/2021
Device Model Number7060-90
Device Lot Number493732
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/04/2018
Initial Report FDA Received Date09/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age37 YR
-
-