• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYNOSURE INC. SCULPSURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYNOSURE INC. SCULPSURE Back to Search Results
Model Number 100-7026-010
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Patient Problem/Medical Problem (2688)
Date of Event 08/26/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
In this incident, patient experienced a blister on the flanks area following a laser procedure with the sculpsure.User error was involved in this incident since it was also determined that the treatment was performed over stretchmarks.The clinical reference guide was not followed as it cautions: "never treat over open areas, rashes, healing wounds or any skin that is not intact.Document any stretch marks, scars, prominent vessels, moles, tattoos, or skin irregularities." the patient sought medical intervention at urgent care.There was no medication prescribed to the patient in this event.The device was evaluated and found to be operating as intended within specification.Blisters are expected side effects from laser treatments, however this is reportable because the patient had medical intervention.
 
Event or Problem Description
Patient had medical intervention following a laser procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCULPSURE
Common Device Name
SCULPSURE
Manufacturer (Section D)
CYNOSURE INC.
5 carlisle road
westford MA 01886
MDR Report Key7911676
Report Number1222993-2018-00021
Device Sequence Number1936610
Product Code PKT
UDI-Device Identifier00841494100032
UDI-Public(01)00841494100032(11)170329
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K171992
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device No Information
Device Model Number100-7026-010
Device Catalogue Number105-7026-000
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/28/2018
Initial Report FDA Received Date09/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age30 YR
-
-