• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ATRIUM MEDICAL PROLITE ULTRA MESH; MESH, SURGICAL, POLYMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ATRIUM MEDICAL PROLITE ULTRA MESH; MESH, SURGICAL, POLYMERIC Back to Search Results
Model Number 30802
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A follow up report will be submitted upon completion of the investigation into this event.
 
Event or Problem Description
When the outer packing was opened they found some foreign matter inside the wrap.The matter was faint yellow and at the edge of the mesh.
 
Additional Manufacturer Narrative
Quality control performed an inspection for foreign materials and submitted samples to the analytical chemistry department for analysis.Analytical personnel characterized the foreign material using microscopic imaging for size and shape and chemical characterization using infrared spectroscopy.Originally, gravimetric analysis was not performed for any samples due to the foreign materials being embedded into the samples.However, after ftir analysis determined that the foreign material on the mesh was fish oil, the amount of fish oil on the mesh could be determined by using a modified version of a test procedure that determines the fish oil coating density of coated mesh products.On the basis of the biological safety risk assessment for c-qur mesh, 0.072 mg of fish oil is not expected to elicit a toxicological risk to the patient.Therefore, the resulting risk of this foreign material is considered to be negligible.Clinical evaluation: it was reported that foreign matter was seen inside the package wrap.Prolite ultra is intended for use in soft tissue deficiencies including hernia repair, traumatic or surgical wounds and chest wall reconstruction procedures requiring reinforcement with a non-absorbable supportive material.When a product is brought to the surgical arena and prepared for use it is inspected to insure the integrity of the packaging, the expiration dates and confirmation of the correct product size and type, etc.Is conducted between two clinicians in accordance with the safe practice recommendations provided by the joint commission.If any discrepancies are discovered the product is rejected and a new one is prepared.This may represent a delay prior to patient contact while the second product is prepared.When this step is omitted, and a product within a defective package is used it would put the patient at increased risk for infection.Any surgery that causes a break in the skin can lead to a postoperative infection.Microorganisms can infect a surgical wound through various forms of contact, such as from the touch of a contaminated caregiver or surgical instrument, through microorganisms in the air, or through microorganisms that are already on or in your body and then spread into the wound.If particulate is introduced into an operative site the patient may experience a foreign body or host tissue reaction, infection, pain, and/or allergic reaction or toxicity.Foreign material that is embedded in the product may cause procedure delay if noticed during set-up.If the product is used the particulate is not expected to illicit a biological response and therefore the risk is negligible.The instructions for use (ifu) states complications that may occur with the use of any surgical mesh include, but are not limited to, pain, inflammation, infection, allergic reaction, seroma, hematoma, fistula formation or mechanical disruption of the tissue and/or mesh material and possibly adhesions when placed in direct contact with the viscera and other organs.The patient should be advised to contact the physician should an adverse reaction occur.The ifu also states that this device is supplied sterile.Please inspect packaging to ensure it is intact and not damaged prior to use.Do not use if package is damaged or opened.
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROLITE ULTRA MESH
Common Device Name
MESH, SURGICAL, POLYMERIC
Manufacturer (Section D)
ATRIUM MEDICAL
40 continental blvd
merrimack NH 03054
MDR Report Key7911891
Report Number3011175548-2018-01035
Device Sequence Number19194834
Product Code FTL
UDI-Device Identifier00650862308020
UDI-Public00650862308020
Combination Product (Y/N)N
Initial Reporter CountryCN
PMA/510(K) Number
K002093
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/29/2022
Device Model Number30802
Device Catalogue Number30802
Device Lot Number424151
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/18/2018
Supplement Date Received by Manufacturer10/30/2018
Initial Report FDA Received Date09/26/2018
Supplement Report FDA Received Date10/31/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-