| Model Number |
989605414121 |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Evaluation of the transducer will be included in a follow up report upon its return and investigation completion.
|
| |
|
Event or Problem Description
|
|
A customer reported an articulation issue with their x7-2t model transducer.There was no injury associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation of the x7-2t model transducer revealed corrosion on the connector and between the knobs on the midhandle.Performance testing of the probe verified proper function of the articulation knobs, so the reported problem could not be confirmed.However, signs of corrosion or material buildup were identified within the knob/midhandle interface which may have led to the issue observed in the field.The investigation concluded that exposure to fluid and materials over time likely resulted in the buildup and reported problem.The damage found on the probe is indicative of improper transducer care and maintenance.
|
| |
|