Model Number 840
Medical Device Problem Codes No Flow (2991); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/03/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
At this time, medtronic has not received the suspect device/component from the customer for evaluation.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that while in use on a patient, the 840 ventilator stopped ventilation and the systems locked out.There was no injury to the patient as a result of the event.
 
Additional Manufacturer Narrative
Device evaluation: it was reported that the unit passed self-testing.The unit however, was noted to not have yet been returned to use.Information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The patient was placed on an alternate ventilator and there was no injury to the patient as a result of the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
840 VENTILATOR
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
COVIDIEN
micheal collins rd mervue
galway
MDR Report Key7912672
Report Number8020893-2018-00472
Device Sequence Number2052385
Product Code CBK
UDI-Device Identifier10884521076976
UDI-Public10884521076976
Combination Product (Y/N)N
Initial Reporter CountryLE
PMA/510(K) Number
K151252
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number840
Device Catalogue Number4-840220AIQC-EN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/06/2018
Supplement Date Received by Manufacturer01/25/2019
Initial Report FDA Received Date09/26/2018
Supplement Report FDA Received Date02/04/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-