| Model Number |
6947M97 |
| Medical Device Problem Codes |
Fracture (1260); Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574)
|
| Health Effect - Clinical Code |
Shock from Patient Lead(s) (3162)
|
| Date of Event |
09/01/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient called to report hearing a lead integrity alert (lia) on the right ventricular (rv) lead and then the ¿lead broke¿ and shocked the patient.A follow up with the patient¿s healthcare provider yielded that the rv lead had fractured and the patient was inappropriately shocked due to oversensing on the rv lead.The lead was capped and a new lead was implanted.No further patient complications have been reported as a result of this event.
|
| |
|