Catalog Number RMT261218
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 09/06/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitant products: pxc121200/11473694 - udi: (b)(4); pxc201400/11264905 - udi: (b)(4); pxc201200/11068406 - udi: (b)(4).The review of the manufacturing paperwork verified that this lot met all pre-release specifications.Per the gore® excluder® aaa endoprosthesis instructions for use (ifu), adverse events that may occur and / or require intervention include, but are not limited to endoleak and aneurysm enlargement.
 
Event or Problem Description
The following was reported to gore: on (b)(6) 2013, the patient presented with an abdominal aortic aneurysm that was treated with the implantation of gore® excluder® aaa endoprostheses.On (b)(6) 2018, a follow-up ct showed endoleak of undetermined origin and a significant aneurysm growth from 75mm (measured on (b)(6) 2017 via ultrasound) to 95mm.No reintervention is currently planned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AORTIC EXCLUDER AAA ENDOPROSTHESIS (C3)
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W.L. GORE & ASSOCIATES
flagstaff AZ
MDR Report Key7913780
Report Number2017233-2018-00576
Device Sequence Number941169
Product Code MIH
UDI-Device Identifier00733132610617
UDI-Public00733132610617
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Reporter Type Manufacturer
Report Source study
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/29/2016
Device Catalogue NumberRMT261218
Device Lot Number11219174
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date09/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age87 YR
Patient Weight85
-
-