| Model Number |
N/A |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/13/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device was not implanted.The actual device has been returned for evaluation.The investigation is currently ongoing.A follow up report will be submitted once the investigation is complete.A review of the device history record of the product code/lot# combination was conducted with no findings.
|
| |
|
Event or Problem Description
|
|
The user facility reported the ik sheath was kinked out of the package.The product never touched the patient.The procedure completed successfully with a second sheath.There was no patient injury or medical intervention required, the patient was in stable condition.Additional information was received on september 17, 2018.The procedure being performed was a peripheral intervention.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted as follow up no.1 to provide the completed investigation.One 5fr ik sheath and dilator were received for product evaluation.The device was returned out of the package and a part of the label was returned separately.Visual inspection revealed that there were no gross visual anomalies noted with the dilator.The sheath was examined, and there was a kink in the sheath near the hub.There is no evidence that this event was related to a device defect or malfunction.The exact cause of the reported event cannot be definitively determined based on the available information.
|
| |
|