MAUDE Adverse Event Report: JOHNSON AND JOHNSON SURGICAL VISION INC. TECNIS; TORIC IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON AND JOHNSON SURGICAL VISION INC. TECNIS; TORIC IOLS Back to Search Results
Model Number ZCT225
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); No Code Available (3191)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: unknown, not provided, but the best estimate date is during 2018.Unknown if the lens was implanted.Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to johnson and johnson vision has been submitted.
 
Event or Problem Description
It was reported that intraocular lens (iol) was defective.No further information was provided.
 
Additional Manufacturer Narrative
Additional information: additional information was received and it was learnt that the lens was explanted because of patient experiencing refractive error.Patient had recovered.No further information received.If implanted, give date: (b)(6) 2018.If explanted, give date: (b)(6) 2018.Device available for evaluation? yes.Returned to manufacturer on: 10/1/2018.Device returned to manufacturer? yes.Patient code 3191 ¿ explant.Patient code 3191 ¿ refractive error.Device evaluation: the product was received in a biohazard bag.The lens was placed in a sample bottle and product was found cut in pieces, most probably to make explant or removal and replacement possible.Considering the condition of the lens additional analysis is not possible.The complaint cannot be confirmed.Manufacturing record review: the manufacturing process record was evaluated and no deviation was found during process related to the complaint issue reported.There was no discrepancy found during the mrr (manufacturing record review).The product was manufactured and released according to specifications.The search revealed no other complaints have been received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Common Device Name
TORIC IOLS
Manufacturer (Section D)
JOHNSON AND JOHNSON SURGICAL VISION INC.
santa ana CA
MDR Report Key7913980
Report Number3011852734-2018-00167
Device Sequence Number19168376
Product Code HQL
UDI-Device Identifier05050474552470
UDI-Public(01)05050474552470(17)220222
Combination Product (Y/N)Y
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source distributor,health profession
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/03/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/22/2022
Device Model NumberZCT225
Device Catalogue NumberZCT225U070
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/01/2018
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 08/28/2018
Supplement Date Received by Manufacturer10/03/2018
10/25/2020
Initial Report FDA Received Date09/27/2018
Supplement Report FDA Received Date10/26/2018
11/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-