COOK INC APPROACH CTO MICROWIRE WIRE GUIDE; DQX WIRE, GUIDE, CATHETER Back to Search Results
Model Number G50793
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/29/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The field was not populated due to a technical limitation.Product code: dqx.This mdr is being filed after the associated complaint was reviewed under remediation protocol (b)(4), complaint/mdr retrospective review and remediation and reassessed as reportable.Additional complaint investigation and record remediation was not performed.
 
Event or Problem Description
It was reported that while the approach cto microwire wire guide was being advanced into the anterior tibial artery (ata), it stopped advancing and became stuck within the catheter.It could not be removed alone, so the wire and catheter were removed together, and the procedure was completed with another set of devices.It was later found that the wire was damaged and partially unraveled near the distal tip.There were no injuries or additional procedures reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APPROACH CTO MICROWIRE WIRE GUIDE
Common Device Name
DQX WIRE, GUIDE, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7914477
Report Number1820334-2018-03036
Device Sequence Number16540559
Product Code DQX
UDI-Device Identifier00827002507937
UDI-Public(01)00827002507937(17)190801(10)5796306
Combination Product (Y/N)N
Initial Reporter CountryMY
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/01/2019
Device Model NumberG50793
Device Catalogue NumberCMW-14-300-25G
Device Lot Number5796306
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2016
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/13/2018
Initial Report FDA Received Date09/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age69 YR
-
-