| Catalog Number |
367964 |
| Medical Device Problem Codes |
Short Fill (1575); Volume Accuracy Problem (1675)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/04/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown.The date received by manufacturer has been used for this field.Multiple lot numbers: there were multiple lot numbers reported to be involved.The information for each lot number is as follows: medical device lot #: 7305700, medical device expiration date: 2018-11-30, device manufacture date: 2017-11-01.Medical device lot #: 7340962, medical device expiration date: 2018-12-31, device manufacture date: 2017-12-06.A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that bd vacutainer® pst¿ gel and lithium heparinn (lh) blood collection tubes under filled.No serious injury or medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: bd had not received samples or photos from the customer facility for evaluation; therefore, the investigation was limited.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Bd has initiated further investigation relating to this issue through a capa.The investigation is still on-going and improvements will be made as the potential causes of this issue are identified.As no samples or photos were received for evaluation, the customer's indicated failure mode was not observed by bd.However, further investigation has been initiated through a capa.The investigation is still on-going and improvements will be made as the potential causes are identified.Although no samples or photos were available for evaluation, bd has initiated further investigation through a capa to identify the potential root cause(s).
|
| |
|
Event or Problem Description
|
|
It was reported that bd vacutainer pst gel and lithium heparinn (lh) blood collection tubes under filled.No serious injury or medical intervention was reported.
|
| |
|
Search Alerts/Recalls
|