SENDX MEDICAL, INC. ABL80 FLEX ANALYZER; ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH Back to Search Results
Model Number 393-839
Medical Device Problem Codes Contamination (1120); Fracture (1260); Contamination of Device Ingredient or Reagent (2901)
Health Effect - Clinical Code Abnormal Blood Gases (1034)
Date of Event 09/10/2018
Type of Reportable Event Serious Injury
Event or Problem Description
A patient analysis was performed on the subject analyzer, abl80 flex serial number (b)(4), and based on the reported hct/thb result the patient received a blood transfusion.The user was suspicious of the subject analyzer's result and therefore performed another measurement on a comparison abl80 flex that was on-site.Measured results: subject analyzer: thb = 7.1g/dl hct = 22%.Comparison analyzer: thb = 13.7 g/dl hct = 42%.Based on a comparison of the results between the subject and comparison analyzers, the user concluded that the subject analyzer results were incorrect and the blood transfusion was unnecessary.A service technician evaluated the analyzer and found that the roller wheel assembly was not working properly.Further investigation revealed the roller wheel assembly housing was cracked around a set screw that is used to secure the roller wheel assembly to the roller motor shaft.The cracked housing resulted in roller wheel assembly slippage affecting sample aspiration and flushing.The issue was resolved following replacement of the roller wheel assembly.
 
Event or Problem Description
A patient analysis was performed on the subject analyzer, abl80 flex serial number (b)(4)., and based on the reported hct/thb result the patient received a blood transfusion.The user was suspicious of the subject analyzer's result and therefore performed another measurement on a comparison abl80 flex that was on-site.Measured results: subject analyzer: thb = 7.1g/dl hct = 22%.Comparison analyzer: thb = 13.7 g/dl hct = 42%.Based on a comparison of the results between the subject and comparison analyzers, the user concluded that the subject analyzer results were incorrect and the blood transfusion was unnecessary.A service technician evaluated the analyzer and found that the roller wheel assembly was not working properly.Further investigation revealed the roller wheel assembly housing was cracked around a set screw that is used to secure the roller wheel assembly to the roller motor shaft.The cracked housing resulted in roller wheel assembly slippage affecting sample aspiration and flushing.The issue was resolved following replacement of the roller wheel assembly.Update on 04/30/19.Device history record (dhr) for abl80 with serial number (b)(4) was reviewed.The review did not reveal any discrepancies or failures.Dhr sensor cassette sc80 with serial number (b)(4) was reviewed.There were no reworks or non-conformance associated with this sensor cassette.Both dhrs indicated that the products met all of the required specifications.The date of the incident is 09/10/18 @13:58.The roller wheel was replaced by an field service engineer on 09/27/18.The analyzer logs contain records from 04/28/18 to the rolloe wheel replacedment date of 09/27/18 was reviewed.The calibration logs do not show evidence calibration errors that would be consistent with a malfunctioning roller wheel.The analyzer's manual quality control (mqc) logs reveal that the end-user does not perform routine quality control with radiometer qc material as recommened.The end-user did perform qc on 09/11/18 ( the day following the incident) and the hct results were within performance speciifcations.The analyzer logs do not show clear evidence that the crach in the roller wheel affected analyzer performance leading up to and following the incident date.Since ethe roller wheel could not be returned for confirmation testing, this investigation into the roller wheel will conclude as "unable to confirm" with high confidence that the roller wheel did not contribute to the reported (-) hct bias.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABL80 FLEX ANALYZER
Common Device Name
ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH
Manufacturer (Section D)
SENDX MEDICAL, INC.
1945 palmoar oaks way
carlsbad CA 92011
MDR Report Key7914865
Report Number2027541-2018-00002
Device Sequence Number2255058
Product Code JPI
UDI-Device Identifier05700693938394
UDI-Public(01)05700693938394
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K051804
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number393-839
Device Catalogue Number393-839
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/10/2018
Supplement Date Received by Manufacturer04/10/2019
Initial Report FDA Received Date09/27/2018
Supplement Report FDA Received Date05/01/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-