Catalog Number 466P306AU
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Embolus (1830); Occlusion (1984); Perforation (2001)
Event Date 08/08/2017
Event Type  Injury  
Manufacturer Narrative
Please note that the exact event date is unknown.The product was not returned for analysis and the sterile lot number has not been provided; therefore, no device analysis nor device history record review could be performed.Additional information is pending and will be submitted within 30 days of receipt.
 
Event Description
As reported by the legal brief, the patient underwent placement of a trapease vena cava filter.The report states that the filter subsequently malfunctioned and caused injury and damage to the patient including, but not limited to perforation of the inferior vena cava filter struts.As a direct and proximate result of these malfunctions, the patient suffered life-threatening injuries and damages, and required extensive medical care and treatment.As a further proximate result, the patient has suffered and will continue to suffer significant medical expenses, and pain and suffering, and other damages.
 
Manufacturer Narrative
After further review of additional information received, the following sections have been updated accordingly.As reported, the patient underwent placement of a trapease inferior vena cava (ivc) filter.The indication for filter placement is not available.The report states that the filter subsequently malfunctioned and caused injury and damage to the patient including, but not limited to, perforation of the inferior vena cava filter struts.The filter remains implanted; thus, unavailable for analysis.The product was not returned for analysis and the sterile lot number has not been provided; therefore, no device analysis nor device history record review could be performed.The trapease vena cava filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the vena cava for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pulmonary embolism where anticoagulant therapy has failed, or is contraindicated.The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart.It was reported that there was perforation of the ivc; however, a clinical conclusion could not be determined as to the cause of the event.A review of the instructions for use notes vessel damage such as intimal tears and perforation as procedural complications related it ivc filters.Ivc perforation from removable filters is relatively common, and directly related to how long the filter has been in place.Studies have noted a greater than 80% perforation rate overall, with all filters imaged after 71 days from implantation revealing some level of perforation.Clinical factors that may have influenced the event include patient, pharmacological and lesion characteristics.Without procedural films or images for review the reported event(s) could not be confirmed.Given the limited information available for review at this time, there is nothing to suggest that the reported events are related to the design and manufacturing process of the device; therefore, no corrective action will be taken.Should additional information become available, the file will be updated accordingly.
 
Manufacturer Narrative
After further review of additional information received, the following sections have been updated accordingly.Additional information received per the medical records indicate that the patient has a history of deep vein thrombosis and pulmonary embolism.The filter was deployed into the inferior vena cava at the level of the renal veins.  additional information was received per the patient profile form (ppf).It states that results of computed tomography (ct) scans reveal that the patient experienced perforation of filter strut(s) outside the inferior vena cava, blood clots, clotting, and/or occlusion of the inferior vena cava.The patient became aware of the reported events approximately six years and ten months after the index procedure.The patient continues to experience emotional distress, mental anguish, anxiety and stress.As reported the patient underwent placement of a trapease inferior vena cava (ivc) filter.Per the medical records, the patient has a history of deep vein thrombosis and pulmonary embolism.The filter was deployed into the inferior vena cava at the level of the renal veins.The report states that the filter subsequently malfunctioned and caused injury and damage to the patient including, but not limited to perforation of the inferior vena cava filter struts.Per the patient profile form (ppf), the results of computed tomography (ct) scans reveal that the patient experienced perforation of filter strut(s) outside the inferior vena cava, blood clots, clotting, and/or occlusion of the inferior vena cava.The patient reports emotional distress, mental anguish, anxiety and stress.The filter remains implanted; thus, unavailable for analysis.The product was not returned for analysis.A review of the device history record revealed no anomalies during the manufacturing and inspection processes that can be associated with the reported complaint.The trapease vena cava filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the vena cava for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pulmonary embolism where anticoagulant therapy has failed, or is contraindicated.The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart.It was reported that there was perforation of the ivc; however, a clinical conclusion could not be determined as to the cause of the event.A review of the instructions for use notes vessel damage such as intimal tears and perforation as procedural complications related it ivc filters.Ivc perforation from removable filters is relatively common, and directly related to how long the filter has been in place.Studies have noted a greater than 80% perforation rate overall, with all filters imaged after 71 days from implantation revealing some level of perforation.Blood clots and occlusive thrombosis within the filter and vasculature do not represent a device malfunction.Anxiety does not represent a device malfunction and may be related to underlying patient related issues.Clinical factors that may have influenced the event include patient, pharmacological and lesion characteristics.Without procedural films or images for review the reported event(s) could not be confirmed.Given the limited information available for review at this time, there is nothing to suggest that the reported events are related to the design and manufacturing process of the device; therefore, no corrective action will be taken.Should additional information become available, the file will be updated accordingly.
 
Manufacturer Narrative
Correction to (product code).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAPEASE PVCF FEM/JUG 55CM CSI
Type of Device
FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
CORDIS CASHEL
cahir road
cashel co. tipperary
EI 
MDR Report Key7915033
MDR Text Key121912803
Report Number1016427-2018-01879
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
PMA/PMN Number
K020316
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup,Followup
Report Date 05/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/31/2013
Device Catalogue Number466P306AU
Device Lot Number15166700
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 08/31/2018
Initial Date FDA Received09/27/2018
Supplement Dates Manufacturer Received09/28/2018
11/09/2018
11/09/2018
Supplement Dates FDA Received10/25/2018
12/05/2018
05/28/2019
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Other;
Patient Age62 YR
-
-