WRIGHTS LANE SYNTHES USA PRODUCTS LLC CAM LOCK LEVER FOR RADIOLUCENT AIMING ARMS; GUIDE Back to Search Results
Catalog Number 03.010.497
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/31/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device is an instrument and is not implanted/explanted.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2018, the cam lock lever for radiolucent aiming arm was broken off upon inspection when the femoral recon nail set was opened from the packaging.There was no patient involvement.Concomitant device reported: radiolucent aiming arm (part# 03.033.003, lot# unknown, quantity# 1).This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
The device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.A manufacturing investigation was performed on the complaint device at the manufacturing site and is summarized below.Manufacturing investigation: visual inspection: the shipped device 03.033.003, ¿aim-arm radioluc f/greater trochanter fe¿ with lot number l699665 is in proper condition.No special or heavily used look.The complained component 03.010.497 ¿cam-lock lever f/aiming arm¿ which was built in is broken and was not receipt with the device.Only the rest of the article 03.010.497 ¿cam-lock lever f/aiming arms¿ is left.Functional inspection: since the component 03.010.497 has been broken off, the function test cannot be carried out.Dimensional inspection: since the component 03.010.497 has been broken off, so a dimensional inspection cannot be carried out.Dhr review: the cam lever 03.010.497, is a sub-component of part #: 03.033.003 lot # l699665 , part: 03.033.003, lot: l699665, manufacturing site: haegendorf, release to warehouse date: 22.Feb.2018.The device history record shows this lot was processed through the normal manufacturing and inspection operations with no rework or nonconformities noted.This lot met all dimensional, visual and functional criteria at the time of release with no issues documented during the manufacturing process.Review of the device history record showed that there were no issues during the manufacture of this product which would contribute to this complaint condition.Material/hardness review: the raw material certificate was reviewed and the used raw material ¿peek-classix white¿ (charge # er7650996) was according to the specification of the device.Manufacturing summary/conclusion: the manufacturing evaluation conducted shows that there was no issue during the manufacture of the product that would contribute to this complaint condition.No manufacturing issue was found during investigation.During this investigation no product design or manufacturing issues were observed that may have contributed to the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device history lot the cam lever 03.010.497, is a sub-component of part #: 03.033.003 lot # l699665 , part: 03.033.003, lot: l699665, manufacturing site: haegendorf, release to warehouse date: 22.Feb.2018.The device history record shows this lot was processed through the normal manufacturing and inspection operations with no rework or nonconformities noted.This lot met all dimensional, visual and functional criteria at the time of release with no issues documented during the manufacturing process.Review of the device history record showed that there were no issues during the manufacture of this product which would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAM LOCK LEVER FOR RADIOLUCENT AIMING ARMS
Common Device Name
GUIDE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7915054
Report Number2939274-2018-53967
Device Sequence Number857446
Product Code FZX
UDI-Device Identifier10886982069146
UDI-Public(01)10886982069146
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 09/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.010.497
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/01/2018
Initial Date Received by Manufacturer 09/04/2018
Supplement Date Received by Manufacturer10/01/2018
11/01/2018
01/03/2019
Initial Report FDA Received Date09/27/2018
Supplement Report FDA Received Date10/16/2018
11/01/2018
01/04/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-