• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC. DIMENSION VISTA®; DIMENSION VISTA® LIPL LIPASE FLEX® REAGENT CARTRIDGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC. DIMENSION VISTA®; DIMENSION VISTA® LIPL LIPASE FLEX® REAGENT CARTRIDGE Back to Search Results
Catalog Number K3056 SMN 10461745
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/13/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer contacted the siemens customer care center (ccc) for the discordant depressed lipl results.Siemens headquarters support center (hsc) has reviewed the information provided by the customer.The customer stated that a new flex was loaded onto the instrument and qc was within range.The event is isolated to one flex reagent cartridge of lot 18040ba.Using a new flex resolved the issue.No error flags were generated at the time of testing.Hsc is not able to determine the cause of the low recovery with this flex.Instrument data is not available.There is no evidence of a product non-conformance.The system is working as specified with the new flex.The cause is unknown.The device is operating within specifications.No further evaluation of the device is required.
 
Event or Problem Description
Discordant falsely depressed lipase (lipl) results were obtained on qc and patient samples on the dimension vista 500 system.The results were reported to the physician(s).The patient samples were reprocessed due to qc being out of laboratory ranges.Higher results and within range qc values were obtained on the same samples on an alternate dimension vista and on reprocessing on the original dimension vista with a new flex reagent cartridge from the same lot.Corrected patient results were reported.There are no reports of treatment change or adverse health consequences to the patients due to the discordant falsely depressed lipl results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIMENSION VISTA®
Common Device Name
DIMENSION VISTA® LIPL LIPASE FLEX® REAGENT CARTRIDGE
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
500 gbc drive
po box 6101
newark DE 19714 6101
MDR Report Key7915178
Report Number2517506-2018-00546
Device Sequence Number19067690
Product Code CHI
UDI-Device Identifier00842768020223
UDI-Public00842768020223
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/06/2018
Device Catalogue NumberK3056 SMN 10461745
Device Lot Number18040BA
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/13/2018
Initial Report FDA Received Date09/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-