| Catalog Number |
05168589190 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/21/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This event occurred in (b)(6).
|
| |
|
Event or Problem Description
|
|
The customer stated that they received erroneous results for one patient sample tested with crep2 creatinine plus ver.2 on a cobas 8000 c 702 module.The results did not correspond to the clinical picture of the patient.The sample was initially tested on another c 702 analyzer, resulting with a crep2 value of -4 umol/l.The sample was repeated on the complained c 702 analyzer, resulting with a value of 2 umol/l.The sample was repeated on the complained c 702 analyzer, resulting as 3 umol/l.The results were reported to the doctor who did not believe the results fit the patient's clinical picture.The patient normally has crep2 values of 35 to 45 umol/l.No adverse events were alleged to have occurred with the patient.The c 702 analyzer serial number is (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation found the results are consistent with acetaminophen/paracetamol intoxication.According to product labeling for the assay: acetaminophen intoxications are frequently treated with n-acetylcysteine.N-acetylcysteine at a plasma concentration above 333 mg/l and the acetaminophen metabolite n-acetyl-p-benzoquinone imine (napqi) independently may cause falsely low results.
|
| |
|