• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH AND NEPHEW, INC. R3 36MM ID INTL DLT CER LNR 56MM; PROSTHESIS, HIP, HEMI-, TRUNNION-BEARING, FEMORAL, METAL/POLYACETAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH AND NEPHEW, INC. R3 36MM ID INTL DLT CER LNR 56MM; PROSTHESIS, HIP, HEMI-, TRUNNION-BEARING, FEMORAL, METAL/POLYACETAL Back to Search Results
Catalog Number 71331756
Health Effect - Clinical Codes Unspecified Infection (1930); Pain (1994)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient was complaining of pain.A revision surgery was performed due to infection.
 
Event or Problem Description
It was reported that a revision surgery was performed due to inflammatory pseudotumor after ceramic-on-ceramic total hip.
 
Additional Manufacturer Narrative
The associated complaint devices were returned and evaluated.The lab analysis concluded, minor scratches were seen on the inner surface of the cup, which was likely due to removal.The porous outer surface of the cup had bony growth indicating good fixation within the acetabulum.Bony growth was seen on the porous coated section of the stem, which indicates that the stem had proper fixation within the femur.Corrosion was observed in the mating area of the stem and on the mating region of the modular neck.Edax analysis was used on a smooth section of the neck and a discolored section of the neck to confirm corrosion.The discolored section of the neck resulted in a high oxygen peak on the spectrum and a 23.73% oxygen concentration.This determined that the discoloration was corrosion especially when compared to the smooth section which had a low oxygen peak and an oxygen concentration of only 1.20%.No manufacturing or material deviations were observed during this investigation.A review of the production documentation for the corresponding products did not reveal any deviation from the standard manufacturing processes.A review of complaint history for the listed part revealed no prior complaints for the listed batch.The clinical/medical team concluded, the inflammatory pseudotumor shows similarities with histopathological findings of adverse local tissue reactions.It was suggested that this is related to the ceramic-ceramic pairing.The mentioned publication reports on an inflammatory pseudotumor/adverse local tissue reaction in the absence of cobalt-chromium modular necks as well as on findings similar to reactions known from metal related issues.As the explanted cobalt-chromium femoral neck-stem combination has a known history of metal related failures and the material analysis showed corrosion on the neck taper, it is more likely that the reported findings relate to the modular neck.This was judged in the absence of the ceramic pairing for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 36MM ID INTL DLT CER LNR 56MM
Common Device Name
PROSTHESIS, HIP, HEMI-, TRUNNION-BEARING, FEMORAL, METAL/POLYACETAL
Manufacturer (Section D)
SMITH AND NEPHEW, INC.
1450 brooks road
memphis, tennessee
memphis 38116
MDR Report Key7915434
Report Number1020279-2018-01912
Device Sequence Number277067
Product Code JDH
UDI-Device Identifier03596010619600
UDI-Public03596010619600
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K080625
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/27/2020
Device Catalogue Number71331756
Device Lot Number10DT41753
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 09/03/2018
Supplement Date Received by Manufacturer09/03/2018
09/03/2018
Initial Report FDA Received Date09/27/2018
Supplement Report FDA Received Date11/16/2018
03/04/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
71331856/11DM05787; 71352112/09LTM0003B; 71352504/11AM05044
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-