WRIGHTS LANE SYNTHES USA PRODUCTS LLC 4.5MM VA-LCP CURVED CONDYLAR PLATE/14HOLE/301MM/RT-STER; IMPLANT, FIXATION DEVICE, CONDYLAR PLATE
|
Back to Search Results |
|
| Model Number |
02.124.414S |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
01/01/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Id - case# reported as csn (b)(4).Patient weight is unknown.Date of event is an unknown date in 2018.Additional procodes: hrs, hwc.Complainant part is not expected to be returned for manufacturer review/investigation.Reporter email address is unknown.The investigation could not be completed; no conclusion could be drawn, as no product was received.A device history record review has been requested.(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2018 the patient underwent a hardware removal of a broken variable angle locking compression plate (va-lcp) condylar plate and ten (10) screws.The hardware was originally implanted on (b)(6) 2018.Patient was revised to a retrograde femoral rod.The patient¿s limb was entirely reduced from the initial surgery and allowed for an intramedullary nail (im).The procedure was successfully completed with no surgical delay.Patient status is stable.The surgeon noted that the patient was undergoing radiation treatment for cancer prior to and after the initial surgery on (b)(6) 2018.He believes that this treatment negatively affected the patient¿s ability to create new bone and heal the fracture.After months of load bearing by the plate, as opposed to load sharing which typically happens when a bone begins to heal, the plate snapped.Concomitant devices: trauma screws (part: unknown, lot: unknown, quantity: 10).This report is for a 4.5mm va-lcp curved condylar plate/14-hole/301mm/right-sterile.This is report 1 of 1 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Part: 02.124.414s; lot: l337638; manufacturing site: (b)(4); release to warehouse date: april 04, 2017; expiry date: march 01, 2027.The device history record shows this lot was processed through the normal manufacturing and inspection operations with no rework or nonconformities noted.This lot met all dimensional, visual and packaging criteria at the time of release with no issues documented during the manufacturing process.The raw material was confirmed to be correct per the specification and norm (b)(4) for stainless steel implants for surgery with no non-conformance noted.Review of the device history record showed that there were no issues during the manufacture of this product which would contribute to this complaint condition.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|