COVIDIEN 840 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 840
Medical Device Problem Code No Flow (2991)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 09/03/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
After the unit was removed from usage, the ventilator was found to not pass the power on self-test (post).A service engineer (se) inspected the device and replaced the graphical user interface (gui) to breath delivery (bd) cable with a provisory one.The unit passed testing and operated within the manufacturing specifications.Once the definitive cable is available, the part will be installed.Should the replaced component be returned to the manufacturing site, an additional evaluation will be performed and a supplemental report will be submitted.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that while in use on a patient, the 840 ventilator stopped ventilating.The patient was placed on an alternate ventilator and there were no significant consequences to the patient as a result of the event.
 
Additional Manufacturer Narrative
Additional information received as follows: after the ventilator was removed from usage, the ventilator was found to not pass the power on self-test (post).The service engineer replaced the breath delivery (bd) central processing unit printed circuit board.Additionally the provisory bd cable was replaced with the definitive cable.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
On 2018-dec-04 it was identified that a correction needed to be made to the device evaluation summary.Corrected "the service engineer replaced the breath delivery (bd) central processing unit printed circuit board.Additionally the provisory bd cable was replaced with the definitive cable" to "a service engineer (se) inspected the device and replaced the breath delivery (bd) central processing unit (cpu) printed circuit board (pcb) and ventilator arm.Additionally, the provisory bd cable was replaced with a definitive cable.The unit passed testing and operated within the manufacturing specifications." if information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Device evaluation summary: the gui to bd cable assembly was returned to medtronic for failure investigation.The component was visually inspected and functionally tested with no malfunction or product deficiency identified.The bd cpu pcb was also returned to medtronic for failure investigation.A visual inspection of the returned component revealed no anomalies.The component was attached to the failure investigation test ventilator for analysis.The ventilator powered up normally with no errors recorded in the diagnostic logs.All pre-use testing was performed successfully.The ventilator was placed in normal ventilation mode and ran for a period of time and then generated a "vent inop" condition.The fault was isolated to the alliance manufactured dram device.Alliance manufactured devices were discontinued and replaced by oki manufactured devices for new bd pcbs from 2005-mar-14 due to higher reliability of the oki dram devices than the alliance parts.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
840 VENTILATOR
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
COVIDIEN
micheal collins rd mervue
galway
MDR Report Key7915914
Report Number8020893-2018-00474
Device Sequence Number19025307
Product Code CBK
UDI-Device Identifier10884521077102
UDI-Public10884521077102
Combination Product (Y/N)N
Initial Reporter CountryPO
PMA/510(K) Number
K151252
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 01/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number840
Device Catalogue Number4-840220DIPC-PR
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/10/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/04/2018
Supplement Date Received by Manufacturer10/15/2018
12/04/2018
01/17/2019
Initial Report FDA Received Date09/27/2018
Supplement Report FDA Received Date10/26/2018
12/05/2018
01/28/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Weight70
-
-