Model Number 760
Medical Device Problem Code No Flow (2991)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 09/06/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Covidien/medtronic has not received the device/component from the customer for evaluation.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during use on patient the 760 ventilator had generated ventilation failure alarm.The device turned on again then post(power on self test) failure occurred with error codes.The patient was removed from the ventilator and placed on an alternate ventilator with no harm or injury.
 
Additional Manufacturer Narrative
Device evaluation summary: the service personnel (sp) inspected the ventilator and replaced the pressure solenoid (p-sol) board.The ventilator passed all test and calibrations at the time of service and was returned to the customer.One psol was returned to medtronic for further analysis.A visual inspection of the returned psol shows no anomalies.The psol was attached to the failure investigation (f.I) test ventilator and it did not pass the power on self test (post).The fault was isolated to component u38.The cause of the observed condition was determined to be the electronic component failure on the p-sol board.The event will be included in trending and monitoring.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
760 VENTILATOR
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
COVIDIEN
micheal collins rd mervue
galway,gw
MDR Report Key7916047
Report Number8020893-2018-00476
Device Sequence Number13667134
Product Code CBK
UDI-Device Identifier10884521180918
UDI-Public10884521180918
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K990897
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 02/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number760
Device Catalogue NumberG-760110DIUU-JP
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/05/2020
Initial Date Received by Manufacturer 09/10/2018
Supplement Date Received by Manufacturer02/28/2020
Initial Report FDA Received Date09/27/2018
Supplement Report FDA Received Date02/28/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-