MAUDE Adverse Event Report: MASIMO - 40 PARKER RAD-5V PULSE OXIMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MASIMO - 40 PARKER RAD-5V PULSE OXIMETER Back to Search Results
Model Number 25199
Medical Device Problem Code Display or Visual Feedback Problem (1184)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/11/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has been returned to the local facility but has not yet been received at the main office for evaluation.Once the device has been returned and investigated, a follow-up report will be submitted.(b)(6).
 
Event or Problem Description
The customer reported that a segment was missing from the display.No consequences or impact to patient were reported.
 
Additional Manufacturer Narrative
The returned device was evaluated.Visual inspection upon receiving revealed no external damages or defects.Power-on test was performed.The device does turn on using batteries.The middle digit of the red led display on the device does not illuminate.The device was verified operationally and functionally.The device was able to obtain readings.The unit was found to visually and audibly alarm during alarm conditions.Internal inspection revealed a poor solder joint on q5 component causing the middle digit to not illuminate.When mechanical pressure was put on q5, the middle digit began illuminating.A service history record review reveals that this unit was in the field for over eight (8) months with no previous reported issues related to this reported event.Initial reporter zip code exceeded the maximum allowable characters, zip code is as follows: (b)(6).
 
Event or Problem Description
The customer reported that a segment was missing from the display.No consequences or impact to patient were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAD-5V PULSE OXIMETER
Common Device Name
OXIMETER
Manufacturer (Section D)
MASIMO - 40 PARKER
40 parker
irvine CA 92618 1604
MDR Report Key7916349
Report Number2031172-2018-00480
Device Sequence Number2255615
Product Code DQA
UDI-Device Identifier00843997005524
UDI-Public00843997005524
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K033998
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Biomedical Engineer
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 09/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number25199
Device Catalogue Number9197
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/14/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 09/11/2018
Supplement Date Received by Manufacturer10/09/2018
Initial Report FDA Received Date09/27/2018
Supplement Report FDA Received Date10/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-