| Model Number |
70900-01 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/11/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product has been requested back for investigation.A follow- up report will be submitted once additional information is obtained.Note: the meter is designed to report readings of 20 mg/dl to 500 mg/dl.It should be noted that this meter does not give numeric value for readings higher than 500 mg/dl.A "hi" display message indicates a reading higher than 500 mg/dl.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Event or Problem Description
|
|
A health care professional reported receiving a high reading from their adc hospital blood glucose meter.Customer reported a high reading of 500 mg/dl which was higher than a lab reading of 121 mg/dl.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported meter was returned and investigated with retained test strips.Visual inspection has been performed on the meter and no issues were observed.The complaint was not confirmed and no new issues were observed.All results were within range specification and no errors were observed during control solution testing.
|
| |
|
Event or Problem Description
|
|
A health care professional reported receiving a high reading from their adc hospital blood glucose meter.Customer reported a high reading of 500 mg/dl which was higher than a lab reading of 121 mg/dl.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported test strips have not been returned.Extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Dhrs (device history review) for the precision strips were reviewed and the dhrs showed the precision strips passed all tests prior to release.Retain testing was performed for the precision strips indicated by the serial number provided by the customer and all units performed within specification.If the product is returned, the case will be re-opened, and a physical investigation will be performed.
|
| |
|
Event or Problem Description
|
|
A health care professional reported receiving a high reading from their (b)(6) hospital blood glucose meter.Customer reported a high reading of 500 mg/dl which was higher than a lab reading of 121 mg/dl.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|