Model Number N/A
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 06/19/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Exemption number e2016018 b.Braun medical, inc.(importer) is submitting this report on behalf of b.Braun melsungen ag (manufacturer).This report has been identified as b.Braun melsungen ag internal report # (b)(4).No sample has been returned for investigation.Without the actual sample (or device), a thorough investigation could not be performed and no specific conclusion can be drawn as to the cause of the reported event.If the sample and/or additional pertinent information becomes available, a follow up report will be submitted.Device history record (dhr): reviewed the dhr and no abnormality observed during in-process and at final control inspection.
 
Event or Problem Description
As reported by the user facility (translation of user facility information by bbm sales organization in colombia: part of the medical device into patient.
 
Additional Manufacturer Narrative
Exemption number e2016018.B.Braun medical, inc.(importer) is submitting this report on behalf of b.Braun melsungen ag (manufacturer).This report has been identified as b.Braun melsungen ag internal report # (b)(4).We received one used capillary part of an introcan safety pur 22g, 0.9x25mm-sa without packaging.The needle, the capillary housing and the protective cap were not handed over by the customer.The returned capillary was taken to a visual inspection.The capillary is cut off approximately 24 mm behind the capillary tip.The surface of the cut and the shape of this area leads to the assumption that this damages is caused by a pair of scissors.Further one, the capillary was penetrated by the cannula approximately 16 mm away from the capillary tip.The area of the puncture shows a v-shaped damage.This puncture was caused by the cannula by withdrawing and pushing forward again.According to both damages, we assume a problem during application.Therefore this complaint is not confirmed.The complained sample was forwarded to our manufacturing site for evaluation: cut off capillary defect most likely not appear to be attributed by manufacturing process as the defect is able to be detected by the in-line vision system.Damages induced after assembly process is not possible since the catheter had been protected with the protective cap.The simulation by cannula reinsertion did not produce similar defect as per complaint sample.The complaint sample defect matches with the simulation sample of being cut by sharp object such as a scissors or scalpel.The defect on the complaint sample is believed to be occurred after the cannulation process due to been cut by a sharp tool.As communicated in ifu, warning section stated that: after withdrawal, do not reintroduce the steel needle into the catheter, as the latter may be cut off, leading to catheter embolism.Do not use scissors or sharp instruments at or near the insertion site.Cause : defect due to wrong handling.
 
Additional Manufacturer Narrative
Exemption number e2016018.B.Braun inc.(bbmi) (importer) is submitting the report on behalf of b.Braun melsungen ag (manufacturer).This report has been identified as b.Braun medical inc.Internal report number (b)(4).This report was filed for item number, 4251628-04, which is not sold in the united states, however a similar item number, 4251628-02, is sold in the united states by b.Braun medical, inc.The 510(k) has been updated to reflect a 510(k) that item 4251628-02 has been cleared under.Note: b.Braun is aware that the listed exemption has been withdrawn, however since the initial mdr report was filed under this exemption, the follow up is also being filed under the exemption.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTROCAN SAFETY®
Common Device Name
CATHETER,INTRAVASCULAR,THE
Manufacturer (Section D)
B. BRAUN MELSUNGEN AG
carl-braun-str. 1
melsungen, hessen 34212
GM  34212
MDR Report Key7916716
Report Number9610825-2018-00204
Device Sequence Number1488869
Product Code FOZ
UDI-Device Identifier04046964399091
UDI-Public(01)04046964399091(17)230101(10)18A31G8913
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K020785
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/03/2019
Report Date (Section F) 07/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/01/2023
Device Model NumberN/A
Device Catalogue Number4251628-04
Device Lot Number18A31G8913
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2018
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date09/27/2018
Device Age8 MO
Event Location Hospital
Date Report to Manufacturer07/03/2019
Initial Date Received by Manufacturer 09/10/2018
Supplement Date Received by Manufacturer09/10/2018
09/10/2018
Initial Report FDA Received Date09/27/2018
Supplement Report FDA Received Date12/14/2018
07/03/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1