| Model Number |
ZCT450 |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/09/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If explanted; give date: n/a (not applicable).The lens remains implanted.The intraocular lens (iol) is not returning for evaluation as lens remains implanted; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, labeling, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that model zct450 tecnis toric iol (intraocular lens) was placed at 160 degrees in the patient's left eye (os) and was measured at 170 degrees in the 1-month study visit.This outcome did not significantly interfere with activities of daily life.There was no patient injury.No intervention has been planned or performed and the lens remains implanted.No additional information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
|
| |
|