• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. SENSOR ENLITE MMT-7008A; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. SENSOR ENLITE MMT-7008A; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-7008A
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/04/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Inspected 1 opened/used sensor and performed visual inspection and found sensor cannula retracted inside sensor base.Also, found sensor cannula damaged, missing broken part.Also, performed visual inspection and checked insertion needle for burrs/hooks and dull needle tip defects and found a dull point at the tip of the insertion needle.Unable to confirm customer received sensor in said condition due to customer returned opened/used.
 
Event or Problem Description
The customer reported via phone call that the sensor had needle pulling out the cannula when removed it, the customer¿s blood glucose was 111 mg/dl.The sensor will be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSOR ENLITE MMT-7008A
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7916912
Report Number2032227-2018-34583
Device Sequence Number770916
Product Code OZO
UDI-Device Identifier20643169646432
UDI-Public(01)20643169646432(017)20180311
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date03/11/2018
Device Model NumberMMT-7008A
Device Catalogue NumberMMT-7008A
Device Lot NumberHG21U21
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/22/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/07/2018
Initial Report FDA Received Date09/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age52 YR
Patient Weight210
-
-