| Model Number |
MMT-7008A |
| Medical Device Problem Code |
Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/04/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Inspected 1 opened/used sensor and performed visual inspection and found sensor cannula retracted inside sensor base.Also, found sensor cannula damaged, missing broken part.Also, performed visual inspection and checked insertion needle for burrs/hooks and dull needle tip defects and found a dull point at the tip of the insertion needle.Unable to confirm customer received sensor in said condition due to customer returned opened/used.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the sensor had needle pulling out the cannula when removed it, the customer¿s blood glucose was 111 mg/dl.The sensor will be returned for analysis.
|
| |
|