MAUDE Adverse Event Report: JOHNSON AND JOHNSON SURGICAL VISION, INC. TECNIS SYMFONY TORIC; MULTIFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON AND JOHNSON SURGICAL VISION, INC. TECNIS SYMFONY TORIC; MULTIFOCAL IOLS Back to Search Results
Model Number ZXT150
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that a zxr00 21.5 diopter intraocular lens (iol) was implanted in the patient's left eye (os) on (b)(6) 2017.It was later explanted on (b)(6) 2018, due to mechanical failure.There was an incision enlargement, but no injury to the patient.Reportedly, the patient is doing good.No additional information was provided.
 
Additional Manufacturer Narrative
In review of the initial mdr, the section were not correct.The following sections have been updated accordingly: both adverse event and product problem should have been selected.Serious injury should have been selected.Device available for evaluation? yes.Returned to manufacturer on: 09/26/2018.Device returned to manufacturer? yes.Device evaluation: visual inspection of the returned lens shows the product is damaged, most probably to make explant possible.Investigation of the return sample and the condition of the return sample do not suggest the damage was introduced during manufacturing.The complaint cannot be confirmed.Manufacturing records review: during the manufacturing record review of the production order for this serial number, no product deficiency was identified.The documentation shows that the production order was manufactured according to specifications.Historical data analysis: a search of complaints revealed no other complaints have been received for this production order number.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY TORIC
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON AND JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key7916985
Report Number9614546-2018-00916
Device Sequence Number941181
Product Code POE
UDI-Device Identifier05050474605237
UDI-Public(01)05050474605237(17)220703
Combination Product (Y/N)Y
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/03/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/03/2022
Device Model NumberZXT150
Device Catalogue NumberZXT150U215
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/26/2018
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 08/29/2018
Supplement Date Received by Manufacturer10/16/2018
10/25/2020
Initial Report FDA Received Date09/27/2018
Supplement Report FDA Received Date11/12/2018
11/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
Date Report Sent to FDA01/01/2005
-
-