• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VOLCANO CORPORATION VISIONS PV .035 DIGITAL IVUS CATHETER; CATHETER, ULTRASOUND, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VOLCANO CORPORATION VISIONS PV .035 DIGITAL IVUS CATHETER; CATHETER, ULTRASOUND, INTRAVASCULAR Back to Search Results
Model Number 81234
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The implant or explant dates are not applicable to this device.
 
Event or Problem Description
This case was reviewed and investigated according to the manufacturer's policy.It was reported during a therapeutic peripheral procedure, the image on the screen was flickering.The manufacture's catheter device was disconnected and reconnected; however, the image was poor.No other device was used to complete the procedure.No patient injury occurred.Visual and microscopic inspection and an image test were performed on the returned device.The distal tip was deformed.There was lifted adhesive at the distal fillet.A portion of adhesive from both the proximal and distal fillet was missing from the deep scratches.The probable cause of the missing adhesive could not be determined through the investigation.Strain, impact, and forces associated with use can affect the integrity of the device.It could not be determined how and when the cause of the failure occurred.The manufacturing documentation for this device was reviewed and the device met all quality and manufacturing release criteria.This product problem is being submitted because adhesive was missing from the device and it could not conclusively be determined when the separation occurred.There is a potential for harm if the malfunction were to recur.
 
Event or Problem Description
This case was reviewed and investigated according to the manufacturer's policy.Internal reference: (b)(4).This follow-up supplemental report #1 is being submitted to report additional investigation findings, to wit: there is no potential for harm if the malfunction were to recur.This complaint was reassessed based on an engineering study that indicated, this family of catheters do not exhibit any adhesive detachment during normal use.There is no evidence of abnormal use conditions, it is likely the adhesive detached post-procedure, during the process of returning the device for analysis.There was no patient injury, no status decline, no adverse event, and no unplanned additional medical or surgical intervention reported.There is no potential for harm if the malfunction were to recur.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to the alleged death or deterioration in the state of the health of any person.
 
Additional Manufacturer Narrative
Block h6: supplemental report #1 indicated that the adhesive likely detached post procedure; however, the conclusion code was not updated from 4315 (cause not established) to 4308 (cause traced to transport/storage).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISIONS PV .035 DIGITAL IVUS CATHETER
Common Device Name
CATHETER, ULTRASOUND, INTRAVASCULAR
Manufacturer (Section D)
VOLCANO CORPORATION
2870 kilgore road
rancho cordova CA 95670
Manufacturer (Section G)
VOLCARICA S.R.L.
coyol free zone&business park
alajuela B37
CS   B37
Manufacturer Contact
melissa pieplow
2870 kilgore road
rancho cordova, CA 95670
9163651925
MDR Report Key7917335
Report Number2939520-2018-00110
Device Sequence Number1399992
Product Code OBJ
UDI-Device Identifier00845225002596
UDI-Public(01)00845225002596(17)191231
Combination Product (Y/N)Y
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K153094
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/09/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2019
Device Model Number81234
Device Catalogue Number400-0200.272
Device Lot Number0259 50125399
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/31/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 08/31/2018
Supplement Date Received by Manufacturer01/14/2005
12/13/2021
Initial Report FDA Received Date09/28/2018
Supplement Report FDA Received Date03/04/2020
12/30/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/31/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
Date Report Sent to FDA01/01/2005
-
-