| Model Number |
85910P |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/02/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Facility declined to provide omitted patient information.No tests/laboratory data available.The implant or explant dates are not applicable to this device.
|
| |
|
Event or Problem Description
|
|
This case was reviewed and investigated according to the manufacturer's policy.It was reported during a therapeutic peripheral procedure, outside body, after use, during insertion in the artery, the wire did not go through the "wire exit port." there is no patient injury.The returned device was inspected and damage was observed.A portion of adhesive was missing on the proximal fillet around the scanner ground pad.Additionally, a deep scratch was present on the distal fillet of the distal tip and some adhesive was also missing from that region.The probable cause of the observed missing adhesive could not be conclusively determined by the investigation.The manufacturing documentation for this device was reviewed and the device met all quality and manufacturing release criteria.This product problem is being submitted because adhesive was missing from the device and it could not conclusively be determined when the separation occurred.There is a potential for harm if the malfunction were to recur.
|
| |
|
Event or Problem Description
|
|
This case was reviewed and investigated according to the manufacturer''s policy.Internal reference: (b)(4).This follow-up supplemental report #1 is being submitted to report additional investigation findings, to wit: there is no potential for harm if the malfunction were to recur.This complaint was reassessed based on an engineering study that indicated, this family of catheters do not exhibit any adhesive detachment during normal use.There is no evidence of abnormal use conditions, it is likely the adhesive detached post-procedure, during the process of returning the device for analysis.There was no patient injury, no status decline, no adverse event, and no unplanned additional medical or surgical intervention reported.There is no potential for harm if the malfunction were to recur.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to the alleged death or deterioration in the state of the health of any person.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6: supplemental report #1 indicated that the adhesive likely detached post procedure; however, the conclusion code was not updated from 4315 (cause not established) to 4308 (cause traced to transport/storage).
|
| |
|