Model Number MMT-722NAS
Medical Device Problem Code Power Problem (3010)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/16/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
The customer reported via phone that the insulin pump had motor error alarm.Customer¿s blood glucose was 217 mg/dl at the time of incident.Drive support cap was normal.Customer was rewinding the pump and get no reservoir.Customer stated that the insulin pump was not working.The insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
Insulin pump alarmed motor error during rewind due to motor encoder signal out of phase.Unable to perform the self-test, unexpected restart error test, displacement test, rewind, basic occlusion test, occlusion test, prime test and excessive no delivery test or verify operating currents and battery out limit alarm due to motor error alarms.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-722NAS
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7917564
Report Number2032227-2018-34887
Device Sequence Number16622778
Product Code OYC
UDI-Device Identifier00643169478558
UDI-Public(01)00643169478558
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-722NAS
Device Catalogue NumberMMT-722NAS
Device Lot NumberB3722NASJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/22/2018
Initial Date Received by Manufacturer 09/16/2018
Supplement Date Received by Manufacturer05/03/2019
Initial Report FDA Received Date09/28/2018
Supplement Report FDA Received Date05/07/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age53 YR
Patient Weight190
-
-