| Model Number |
MMT-722NAS |
| Medical Device Problem Code |
Power Problem (3010)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/16/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone that the insulin pump had motor error alarm.Customer¿s blood glucose was 217 mg/dl at the time of incident.Drive support cap was normal.Customer was rewinding the pump and get no reservoir.Customer stated that the insulin pump was not working.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Insulin pump alarmed motor error during rewind due to motor encoder signal out of phase.Unable to perform the self-test, unexpected restart error test, displacement test, rewind, basic occlusion test, occlusion test, prime test and excessive no delivery test or verify operating currents and battery out limit alarm due to motor error alarms.
|
| |
|