Catalog Number 222233
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/23/2013
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The 3 complaint devices was received and evaluated.Visual observation confirms the distal tip of the anchor is broken, but is held in place by the suture, thus confirming this complaint.There is slight tissue debris lodged between the inserter shaft and anchor on all the 3 devices.This type of anchor breakage at its distal tip is consistent with historical investigations in which it was determined that the root cause was likely a combination of torque and bending while inserting the anchor into the bone and also the bone quality.Other than this possibility, we cannot discern a root cause for this failure.A batch review was conducted to determine if there were any internal processing issues, which would have contributed to the nature of the product complaint.Our results indicate that this batch was processed without incident and therefore there is no internally assignable cause for the reported problem.Further, a review into the mitek complaints system revealed no other complaints of any type for this lot of devices that were released to distribution.At this point in time, no corrective action is required and no further action is warranted.However, mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.(b)(4).This report is being filed from the (b)(4) complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
 
Event or Problem Description
It was reported by our affiliate the anchor broke when screwed into the bone.Three additional anchors were used.Three (3) of the four (4) anchors used broke when screwed into the bone.The procedure was completed with like product.Please see also medwatch reports 1221934-2018-50888 and 1221934-2018-50890.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.5 HEALIX BR ANCHOR W/OCORD
Common Device Name
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key7918083
Report Number1221934-2018-50889
Device Sequence Number2269710
Product Code MAI
UDI-Device Identifier10886705007790
UDI-Public10886705007790
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K150209
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/02/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2014
Device Catalogue Number222233
Device Lot Number3578941
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/10/2013
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/02/2013
Initial Report FDA Received Date09/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-