MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINN SECTOR HA ACET CUP 52MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINN SECTOR HA ACET CUP 52MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS Back to Search Results
Catalog Number 121712052
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Pain (1994); Tissue Damage (2104); Not Applicable (3189)
Date of Event 09/05/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Total hip replacement: (b)(6) hospital.Primary (b)(6) 2016.Patient under went at total hip replacement in (b)(6) 2016, 6 months later, he has unexplained pain in the joint, the surgeon first thought it was infected and treated the patient as per an infection.He had fluid collection, with high metal ions reported.The surgeon then thought it could be metallosis.A stage one revision for infection/metallosis was performed.The surgeon would like the prosthesis investigated to help with the reasoning as to what could event may have caused this issue.He would also like the prosthesis sent back to him, after depuy synthes has concluded their investigations, to send to dr walsh for further analysis.
 
Additional Manufacturer Narrative
Pc-(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Update ad 12 nov 2018.Follow up information received that there was no metalosis, no infection and metals ions were within normal limits.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrativ: product complaint # (b)(4).The returned product was examined by a depuy material scientist.There is nothing outward to indicate product error or contribution to the reported problem.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot = null.Device history batch = null.Device history review =null.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN SECTOR HA ACET CUP 52MM
Common Device Name
PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key7918105
Report Number1818910-2018-70782
Device Sequence Number1449853
Product Code LPH
UDI-Device Identifier10603295009160
UDI-Public10603295009160
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K031495
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 09/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/31/2026
Device Catalogue Number121712052
Device Lot Number740461
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2018
Initial Date Received by Manufacturer 09/04/2018
Supplement Date Received by Manufacturer10/04/2018
11/12/2018
03/05/2019
Initial Report FDA Received Date09/28/2018
Supplement Report FDA Received Date10/18/2018
12/03/2018
03/15/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age35 YR
-
-