• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US S-ROM*SLEEVE PRX ZTT, 16D-SML; S-ROM HIP SYSTEM : HIP FEMORAL SLEEVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US S-ROM*SLEEVE PRX ZTT, 16D-SML; S-ROM HIP SYSTEM : HIP FEMORAL SLEEVE Back to Search Results
Catalog Number 550513
Date of Event 09/05/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Total hip replacement: (b)(6) hospital.Primary (b)(6) 2016.Patient under went at total hip replacement in (b)(6) 2016, 6 months later he has unexplained pain in the joint, the surgeon first thought it was infected and treated the patient as per an infection.He had fluid collection, with high metal ions reported.The surgeon then thought it could be metallosis.A stage one revision for infection/metallosis was performed.The surgeon would like the prosthesis investigated to help with the reasoning as to what could event may have caused this issue.He would also like the prosthesis sent back to him, after depuy synthes has concluded their investigations, to send to dr (b)(6) for further analysis.
 
Additional Manufacturer Narrative
Pc(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Update ad 12 nov 2018.Follow up information received that there was no metallosis, no infection and metals ions were within normal limits.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint #
=
> (b)(4).Investigation summary
=
> the returned product was examined by a depuy material scientist.There is nothing outward to indicate product error or contribution to the reported problem.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
S-ROM HIP SYSTEM : HIP FEMORAL SLEEVE
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key7918114
Report Number1818910-2018-70783
Device Sequence Number11330177
Product Code LPH
UDI-Device Identifier10603295175230
UDI-Public10603295175230
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K934412
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 09/04/2018
1 Device was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number550513
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2018
Initial Date Received by Manufacturer 09/04/2018
Supplement Date Received by Manufacturer10/04/2018
11/12/2018
03/05/2019
Initial Report FDA Received Date09/28/2018
Supplement Report FDA Received Date10/18/2018
12/03/2018
03/15/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
-
-