• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRIVIDIA HEALTH, INC. TRUEMETRIX; BLOOD GLUCOSE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRIVIDIA HEALTH, INC. TRUEMETRIX; BLOOD GLUCOSE SYSTEM Back to Search Results
Model Number TRUEMETRIX
Medical Device Problem Codes Unable to Obtain Readings (1516); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Dizziness (2194)
Date of Event 09/04/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Product not returned for evaluation.No replacement product needed at this time.Complaint resolved by training during the time of the call.Product is working as designed.Most likely underlying root cause: mlc-62-user had poor technique test strip udi# (b)(4).Note: manufacturer contacted customer on (b)(4) 2018 in a follow-up call in order to ensure the customer's condition since the initial call; customer condition improved.No symptoms or medical attention reported since the original call.
 
Event or Problem Description
Consumer reported complaint for e-3 error: used test strip, test strip outside of vial too long, sample on top of test strip.Friend is calling on behalf of the customer.The e-3 error occurred when the test strip was inserted into the meter.At the time of the call, the customer reported feeling dizzy.Customer was using proper testing technique.The test strip was not left out of the vial too long and customer stated he did not leave the vial open.During the call on (b)(6) 2018, a blood test was performed by the customer non-fasting and produced test result of 213 mg/dl using truemetrix meter.The product is stored according to specification in the living room.The test strip lot manufacturer's expiration date is 09/16/2019 and open vial date is (b)(6) 2018.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUEMETRIX
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key7918568
Report Number1000113657-2018-00978
Device Sequence Number1594342
Product Code NBW
UDI-Device Identifier00011822445610
UDI-Public(01)00011822445610
Combination Product (Y/N)N
Initial Reporter StateAL
PMA/510(K) Number
K140100
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Remedial Action Other
Type of Report Initial
Report Date (Section B) 09/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/16/2019
Device Model NumberTRUEMETRIX
Device Catalogue NumberRE4H01-81
Device Lot NumberMV2822
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Event Location Home
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/04/2018
Initial Report FDA Received Date09/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-