Model Number CRPLUS
Medical Device Problem Code Power Problem (3010)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 07/27/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Physio-control evaluated the customer's device and verified the reported issue.Physio-control determined that the cause of the reported issue was that internal hybrid layer capacitor (hlc) battery, designator bt3, on the analog pcb assembly had a broken negative tab weld, leaving it open to the device circuit.This resulted in hlc battery bt3 depleting to zero, which could compromise the device's ability to charge and shock if it were necessary.The device did power on, charge and shock during testing.The customer was provided with a replacement device.
 
Event or Problem Description
The customer contacted physio-control to report that their device had displayed a premature low-battery indicator.There were no reports of patient use associated with the reported issue.Upon evaluation of the device, physio-control observed that the internal hybrid layer capacitor (hlc) batteries had depleted, which could inhibit the device's ability to provide defibrillation therapy, if necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK CR(R) PLUS DEFIBRILLATOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
MDR Report Key7918616
Report Number0003015876-2018-01604
Device Sequence Number16622895
Product Code MKJ
UDI-Device Identifier00885074440330
UDI-Public00885074440330
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K033275
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberCRPLUS
Device Catalogue Number99403-000157
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/24/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/07/2018
Initial Report FDA Received Date09/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-