Concomitant medical products- erbe electrosurgical generator.Investigation evaluation: our laboratory evaluation of the product said to be involved confirmed the report.The snare was returned with the snare head slightly advanced outside the distal end of the sheath.The snare head would not retract back into the sheath by handle manipulation.The handle of the device was manipulated to advance the snare head, and when the snare head was advanced outside the sheath, the snare head was misshaped.The snare head did not form a uniform duck bill.When the snare head was advanced outside the sheath, the snare wire was bent above where the duck bill is normally formed.This created two (2) bends in the snare wire.One (1) that was made to create the duck bill shape, and one (1) that was bent during use; thus creating a misshaped snare head.The misshaped snare head contributed to the retraction difficulty that the user encountered.Further, when the snare head was being retracted into the sheath, the additional bend in the snare wire was preventing the snare head from fully retracting.The snare wire was manipulated with a pair of tweezers and formed back into a duck bill shape with one (1) bend in the distal end of the wire.The snare head smoothly advanced and retracted when manipulating the handle.When the snare head was fully retracted inside the sheath, the distance from the distal end of the snare wire to the distal end of the sheath met specification.The sheath of the snare was disassembled from the handle to evaluate the formed snare wire.It was found during the evaluation that the snare wire cannula that is soldered to the snare wire was not in the correct location.The position of this cannula could have contributed to the snare head becoming misshaped during use.Additionally while reviewing the device, two (2) bends were found in the drive wire that could have also contributed to retraction difficulty.The two (2) bends were at 117.6 cm and 149.4 cm from the base of the handle.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: the investigation determined that the device did not meet manufacturing specification.During the manufacturing process for the forming wire, the cannula was not placed at the correct location prior to being soldered to the forming wire.The operators were retrained on the manufacturing process.A corrective action has been initiated to reduce occurrences for snare head retraction difficulties for asdb devices.The product said to be involved is included in the scope of the corrective actions.Prior to distribution, all acusnare polypectomy snares are subjected to a visual inspection and functional testing to ensure device integrity.Corrective action: a corrective action has been initiated in an effort to reduce occurrences of this nature.This product was manufactured prior to implementation of this corrective action.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
|
During an endoscopic mucosal resection (emr), the physician used a cook acusnare polypectomy snare.The snare would not close all of the way.The snare was tested prior to inserting into the endoscope, and visually appeared to work correctly.The physician was able to resect the tissue using the original snare and cautery [intended procedure] even though it would not close properly.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|