AV-TEMECULA-CT ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Catalog Number B2060-080
Medical Device Problem Code Deflation Problem (1149)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/05/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4): during processing of this complaint, attempts were made to obtain complete event, patient and device information.Evaluation summary: visual and dimensional analysis was performed on the returned device.The reported deflation issue was not able to be confirmed due to the balloon being punctured.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history revealed no other incidents.The investigation was unable to determine a cause for the reported difficulties.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was to treat a lesion located in the proximal right superficial femoral artery with a 99% subtotal occlusion in the proximal right superficial femoral artery.A 5.0x100mm self expanding stent was implanted.A 6.0x80 mm armada 35 was inflated one time to nominal pressure to post dilate.After the first inflation, the armada was unable to be deflated.After multiple attempts were made to deflate the armada 35, it was inflated to rated burst pressure, but that did not resolve the deflation issue.With ultrasound guidance, a needle was inserted and successfully punctured the armada 35 balloon.The deflated armada 35 was removed from the anatomy.There were no adverse patient effects and no clinically significant delay in the procedure.There was no additional information provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7918957
Report Number2024168-2018-07505
Device Sequence Number1346571
Product Code LIT
UDI-Device Identifier08717648155284
UDI-Public08717648155284
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K111899
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial
Report Date (Section B) 09/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2020
Device Catalogue NumberB2060-080
Device Lot Number70209G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/19/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/06/2018
Initial Report FDA Received Date09/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-