Catalog Number 400SMTSTD0830
Medical Device Problem Code Difficult or Delayed Activation (2577)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/30/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report numbers: 3005168196-2018-01925; 3005168196-2018-01926; 3005168196-2018-01928.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure using penumbra smart coils (smart coils) and a smart coil detachment handle (handle).During the procedure, the physician inserted a non-penumbra microcatheter into the target location and attempted to place a smart coil, however, the physician repeatedly advanced and retracted the smart coil to reposition it.Consequently, the physician began to experience resistance while advancing near the distal tip of the microcatheter and, therefore, the smart coil was removed.This then occurred with a second smart coil.The physician then successfully placed a smart coil in the target location, however, was unable to detach the coil using the handle.It was reported that there was an abnormal sound when the physician attempted to detach the smart coil with the handle.The physician therefore used a second handle to detach the smart coil.The procedure was then completed using the second handle and additional smart coils.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7919197
Report Number3005168196-2018-01927
Device Sequence Number1913802
Product Code HCG
UDI-Device Identifier00814548014524
UDI-Public00814548014524
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
K160832
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/31/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/06/2023
Device Catalogue Number400SMTSTD0830
Device Lot NumberF81799
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/31/2018
Initial Report FDA Received Date09/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-