Model Number 9735602
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Patient Involvement (2645)
Event Date 09/12/2018
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.Unique device identifier (udi) is unavailable.No parts have been received by the manufacturer for evaluation.Device manufacture date is unavailable.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that, while outside of a procedure, the portable electromagnetic interface could not be recognized by the navigation system.There was no patient present when this issue was identified.No additional information was provided.
 
Manufacturer Narrative
A medtronic representative went to the site to test the equipment.Testing revealed that the electromagnetic interface system of the navigation system was replaced.The system then passed the system checkout and was found to be fully functional.The unit has not been received by the manufacturer for evaluation.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
The electromagnetic interface system for the navigation system was returned to the manufacturer for analysis.The unit was found to be fully functional with no problem found.The reported event could not be duplicated by medtronic personnel.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION COMPACT NAVIGATION SYSTEM
Type of Device
EAR, NOSE AND THROAT STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
stacy ruemping
7000 central avenue ne rcw215
minneapolis, MN 55432
7635260594
MDR Report Key7919486
MDR Text Key122157360
Report Number1723170-2018-04930
Device Sequence Number1
Product Code PGW
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K153247
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 12/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/28/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9735602
Device Catalogue Number9735602
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received11/13/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-