| Model Number |
518-065 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Material Puncture/Hole (1504); Material Perforation (2205); Device Damaged by Another Device (2915)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/04/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient gender and weight not available from the site.Visual examination of catheter confirmed a hole was present 5mm from the distal tip.Likely cause is the user encountered resistance in the anatomy, the catheter deflected, and the guide-wire continued straight, puncturing the catheter.A couple kinks were found along the catheter as well.
|
| |
|
Event or Problem Description
|
|
During a peripheral vascular surgery utilizing femoral access, the surgeon stated when using the quick-cross support catheter 518-065, they could not go as low in the leg as wished, past the proximal anterior tibial artery.Also, when taking out the quick-cross they noticed a hole in the catheter above the tip.No device remnants were detected to be left in the patient.The case was successfully completed with a balloon and guide-wire.No impact on patient outcome due to this event.This is being reported due to the potential for serious injury with recurrence.
|
| |
|
Search Alerts/Recalls
|