• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MEDICAL PRODUCTS KENDALL SCD; SLEEVE, LIMB, COMPRESSIBLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MEDICAL PRODUCTS KENDALL SCD; SLEEVE, LIMB, COMPRESSIBLE Back to Search Results
Model Number 295250
Medical Device Problem Code Electrical Power Problem (2925)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Submitted september 28, 2018 the unit has been received for evaluation.When additional information is completed a follow-up report will be submitted.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The customer reports that there was a booting failure with an audible alarm.A component on the mainboard was burnt.
 
Additional Manufacturer Narrative
An investigation was performed for the reported customer complaint: ¿the customer reports that there was a booting failure with an audible alarm.A component on the mainboard was burnt.¿ the device history record (dhr) for serial number (b)(4) indicates product and specification requirements were met with no non-conforming product identified relating to this customer report.Product cannot be released unless it passes all quality and conformance requirements.In-process procedures are also in place to prevent nonconforming product in the manufacturing process.This ensures components and finished products meet all quality inspection standards.These controls include, but are not limited to: material verification/certification processes, dimensional specifications, statistical samplings, periodic audits, process inspections, machine maintenance/operation and personnel training and certification.The unit was returned to the service center and an evaluation was performed.Thermal damage to the mainboard was visually confirmed.Not all internal components were observed as burnt.The thermal damage to the mainboard appears to be related to over-current or over voltage.The potential root cause was identified as a faulty component, such as a voltage regulator or resistor.A request for further information was made.However, no additional information was provided.Therefore, a root cause could not be determined.The customer also cancelled the service request.The reported customer complaint was confirmed.A specific root cause could not be determined.A probable root cause was determined to be a faulty component.No corrective or preventive action is planned at this time.This information will be utilized for trending purposes to determine the need for corrective actions.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KENDALL SCD
Common Device Name
SLEEVE, LIMB, COMPRESSIBLE
Manufacturer (Section D)
COVIDIEN MEDICAL PRODUCTS
building 10- no 789 puxing roa
covidien 20111 4
CN  201114
MDR Report Key7920487
Report Number3006451981-2018-00609
Device Sequence Number1346540
Product Code JOW
UDI-Device Identifier10884521129573
UDI-Public10884521129573
Combination Product (Y/N)N
Initial Reporter CountryKS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number295250
Device Catalogue Number295250
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/23/2017
Initial Date Received by Manufacturer 03/02/2018
Supplement Date Received by Manufacturer10/12/2018
Initial Report FDA Received Date09/28/2018
Supplement Report FDA Received Date10/17/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-