| Model Number |
ME1001 |
| Medical Device Problem Codes |
Break (1069); Fluid/Blood Leak (1250); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); Blood Loss (2597)
|
| Date of Event |
08/29/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Although requested, the affected product has not been received.A follow up report will be submitted with investigation results should the devices be received for evaluation.
|
| |
|
Event or Problem Description
|
|
The customer reported a generalized recurring product concern in which the male luer is unable to be disconnected and becomes stuck inside the valve or the female luer that it is connected to.The nursing staff have resorted to attempting to remove the tubing set from the patient while using hemostats only to have the male luer break off.In one patient example, the male luer was stuck inside the valve and broke cleanly off from the inside of the spin collar.The nurse could not remember what medication was infusing at the time, however, does recall that the patient lost a negligible amount of blood due to having to change out the entire set-up all the way down to the patient's iv hub.The patient did not require having a new iv site inserted due to the event.There was no report of patient harm.
|
| |
|
Event or Problem Description
|
|
The customer reported a generalized recurring product concern in which the male luer is unable to be disconnected and becomes stuck inside the valve or the female luer that it is connected to.The nursing staff have resorted to attempting to remove the tubing set from the patient while using hemostats only to have the male luer break off.The customer further states that they do not use alcohol on the male luer when they connect the sets together.The nurse will prime the set with the fluid that will be delivered and then connect the syringe to the set and prime until the fluid exits at the male luer with the duct cap still attached., this is also done for dextrose as well.Before the nurse connects the male luer to an open female luer, the nurse will use a prefilled syringe to place saline into the female luer before connection to ensure that an air bubble is not introduced.In one patient example, the male luer was stuck inside the valve and broke cleanly off from the inside of the spin collar.The nurse could not remember what medication was infusing at the time, however, does recall that the patient lost a negligible amount of blood due to having to change out the entire set-up all the way down to the patient's iv hub.The patient did not require having a new iv site inserted due to the event.There was no report of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The affected product has been received and the investigation is pending.A follow up report will be submitted once the failure investigation has been completed.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer¿s report that the bifuse (misreported by customer as trifuse) male luer was unable to be disconnected and became stuck inside the valve or female luer to which it was connected was confirmed.Visual inspection under magnification of the male luer tip¿s broken surface observed that the break was uneven and jagged with corresponding white stress marks.The nature of this break indicates an external application of force.Functional testing was not performed since visual inspection of the sample confirmed the customer¿s report.The root cause of the customer¿s report was the external application of force to the male luer tip.
|
| |
|
Event or Problem Description
|
|
The customer reported a generalized recurring product concern in which the male luer is unable to be disconnected and became stuck or break inside the valve or the female luer connection site, requiring the nursing staff to use hemostats to remove the set.The customer further states that they do not use alcohol on the male luer when they connect the sets, and prime the set with the fluid or dextrose that will be delivered until the fluid exits at the male luer with the duct cap still attached, and the female luer is also primed with saline.In this example, the male luer was stuck inside the valve and broke cleanly off from the inside of the spin collar.The nurse could not remember what medication was infusing at the time, however, does recall that the patient lost a negligible amount of blood due to having to change out the entire set-up all the way down to the patient's iv hub.The iv was saved.There was no report of patient harm.
|
| |
|
Search Alerts/Recalls
|