| Model Number |
800 |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
Test Result (2695)
|
| Date of Event |
07/27/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
On 27 jul 2018, the field service engineer (fse) was at the customer site and observed that the dxh800 instrument had high platelets at startup.The fse replaced the rbc to resolve the reported event.The assignable cause of the reported event was the rbc bath.(b)(4).
|
| |
|
Event or Problem Description
|
|
The customer reported that platelet (plt) results for patients samples are running erroneously high on their unicel dxh 800 coulter cellular analysis system.There was no change to patient treatment reported as a result of this event.The customer did not provide patient data.
|
| |
|