| Model Number |
AC*T DIFF 2 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/11/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The fse adjusted the calibration factors on the instrument to bring the plt results into the center of the range.The fse ran verification testing on the instrument, which all passed.The bec internal identifier for this event is case-(b)(4).
|
| |
|
Event or Problem Description
|
|
The customer reported issues with platelet (plt) quality control results on a coulter ac·t diff 2 analyzer.The customer called customer technical support and refused to adjust their calibration values as recommended, adn requested service.The field service engineer (fse) evaluated the instrument at the customer's site and observed falsely elevated plt values on patient samples.Erroneous results were not released from the laboratory and there was no change or effect to patient treatment associated with this event.Patient results and patient demographic data were not provided by the customer for this event.
|
| |
|