MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 2.7MM VA LCKNG SCREW SLF-TPNG WITH T8 STARDRIVE RECESS 22MM; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 2.7MM VA LCKNG SCREW SLF-TPNG WITH T8 STARDRIVE RECESS 22MM; SCREW, FIXATION, BONE Back to Search Results
Model Number 02.211.022
Medical Device Problem Codes Fitting Problem (2183); Device Damaged by Another Device (2915); Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient dob & weight is not available for reporting.Date of event is unknown.Additional product code: hrs.Not implanted or explanted, partial device remained in patient.Complainant device is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on an unknown date, the variable angle (va) self tapping locking screw went through the va locking compression plate (lcp) posterolateral distal humerus plate.The surgeon was using the va elbow system.The case was going fine until the last screw placement into the plate.The surgeon pre-drilled with a drill bit on a nominal angle.The surgeon powered the locking screw down, and stopped short of the plate and went to tighten by hand and the screw went directly through plate countersinking into the bone.The screw was not into the joint so the surgeon decided to leave the screw implanted in the bone.No action was done due to it would require taking all of the screws and the plate off and starting over.If the screw migrates into the joint, the patient would have to come back for a revision.It is unknown if there was a surgical delay.Patient and surgical outcome is unknown.Concomitant device reported: drill / drill bit (part# 323.062 lot# unknown, quantity# 1).This complaint involves one (1) device, - variable angle (va) self tapping locking screw.This report is 1 of 2 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.7MM VA LCKNG SCREW SLF-TPNG WITH T8 STARDRIVE RECESS 22MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7920760
Report Number2939274-2018-53988
Device Sequence Number13918090
Product Code HWC
UDI-Device Identifier10886982053626
UDI-Public(01)10886982053626
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K100776
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number02.211.022
Device Catalogue Number02.211.022
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/04/2018
Supplement Date Received by Manufacturer10/24/2018
Initial Report FDA Received Date09/28/2018
Supplement Report FDA Received Date10/24/2018
Patient Sequence Number1
Patient Age79 YR
-
-