• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LEGION PS NP FEM SZ 5 LT; PRSTHSIS,KNEE,PATLLFMOROTIBIAL,SMICONSTRAIND,CMNTD,POLYMER/METL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LEGION PS NP FEM SZ 5 LT; PRSTHSIS,KNEE,PATLLFMOROTIBIAL,SMICONSTRAIND,CMNTD,POLYMER/METL Back to Search Results
Catalog Number 71423225
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Injury (2348)
Date of Event 10/04/2016
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a revision surgery has been performed due to a septic arthritis development.An incision and drainage with irrigation and debridement was performed on the left knee.
 
Additional Manufacturer Narrative
The affected genesis ii oval resurfacing patellar, legion non-porous femoral component, genesis ii tibial baseplate and genesis ii ps high flexion insert were not returned for evaluation.Our clinical analysis noted that no lab results or reports were provided to confirm the reported infection.However, the revision op report indicated there was some slimy granulation and synovium throughout the knee, this was debrided and all of the reported ¿infected tissue¿ was removed.Based on the limited information provided and without the return of the device for evaluation, there is no evidence to support our device contributed to the reported infection and as it was 2 years post op if it was infected it was likely a late hematogenously spread, therefore a thorough medical assessment cannot be completed.Our investigation including a review of the manufacturing records for the listed batches did not reveal any deviation from the standard manufacturing processes.Products were sterilized according to sterilization release documentation from quality control.A review of complaint history for the listed parts revealed no prior complaints for the listed failure mode with the same batch number.Without the return of the actual product involved, our investigation of this report is inconclusive.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEGION PS NP FEM SZ 5 LT
Common Device Name
PRSTHSIS,KNEE,PATLLFMOROTIBIAL,SMICONSTRAIND,CMNTD,POLYMER/METL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key7921367
Report Number1020279-2018-01977
Device Sequence Number11168976
Product Code JWH
UDI-Device Identifier03596010607607
UDI-Public03596010607607
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K060742
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/26/2023
Device Catalogue Number71423225
Device Lot Number13HM19654
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/05/2018
Supplement Date Received by Manufacturer09/05/2018
Initial Report FDA Received Date09/30/2018
Supplement Report FDA Received Date02/05/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age67 YR
-
-