| Catalog Number |
RONYX35022UX |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/24/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
During the procedure an attempt was made to use one resolute onyx rx coronary drug eluting stent to treat a moderately tortuous and calcified lesion located in the distal rca exhibiting 95% stenosis.There were no abnormalities in relation to the anatomy.There was no damage noted to packaging.There were no issues noted when removing the device from the hoop.The device was inspected with no issues noted.Negative prep was performed with no issues noted.The lesion was pre dilated.The device did not pass through a previously-deployed stent.Resistance was encountered when advancing the device.Excessive force was used during delivery.It was reported that the device did not cross due to calcified lesion.The stent was withdrawn into the guide catheter and removed and a strut was observed to be lifted up on the proximal end of the stent.Another onyx stent was used successfully to treat the lesion.No patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis summary: the stent was positioned between the markerbands but not meeting specifications due to deformation at the most proximal end.Deformation was evident to the distal stent wraps with struts raised.Slight deformation was evident to the distal tip.The inner lumen patency was verified with a 0.015 inch mandrel.No other damage evident to the remainder of the device.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|