• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND RESOLUTE ONYX RX; STENT, CORONARY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND RESOLUTE ONYX RX; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Catalog Number RONYX35022UX
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/24/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
During the procedure an attempt was made to use one resolute onyx rx coronary drug eluting stent to treat a moderately tortuous and calcified lesion located in the distal rca exhibiting 95% stenosis.There were no abnormalities in relation to the anatomy.There was no damage noted to packaging.There were no issues noted when removing the device from the hoop.The device was inspected with no issues noted.Negative prep was performed with no issues noted.The lesion was pre dilated.The device did not pass through a previously-deployed stent.Resistance was encountered when advancing the device.Excessive force was used during delivery.It was reported that the device did not cross due to calcified lesion.The stent was withdrawn into the guide catheter and removed and a strut was observed to be lifted up on the proximal end of the stent.Another onyx stent was used successfully to treat the lesion.No patient injury reported.
 
Additional Manufacturer Narrative
Product analysis summary: the stent was positioned between the markerbands but not meeting specifications due to deformation at the most proximal end.Deformation was evident to the distal stent wraps with struts raised.Slight deformation was evident to the distal tip.The inner lumen patency was verified with a 0.015 inch mandrel.No other damage evident to the remainder of the device.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTE ONYX RX
Common Device Name
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key7921740
Report Number9612164-2018-02578
Device Sequence Number1643727
Product Code NIQ
UDI-Device Identifier00643169557093
UDI-Public00643169557093
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P160043
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/06/2020
Device Catalogue NumberRONYX35022UX
Device Lot Number0009024989
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/14/2018
Initial Date Received by Manufacturer 09/24/2018
Supplement Date Received by Manufacturer01/08/2019
Initial Report FDA Received Date10/01/2018
Supplement Report FDA Received Date01/16/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age70 YR
-
-