Evaluation summary: the nanocross elite dilatation catheter was received for evaluation.No ancillary devices or cine images from the procedure were received.The nanocross elite balloon chamber was received in a post-inflation profile with fluid in the balloon chamber.Stretching could be seen on the inner guidewire lumen within the balloon chamber.Examination of the proximal balloon bond revealed a crimp of the outer shaft of the inflation lumen.A 0.035¿ guidewire was loaded through, both, the distal tip and the y-manifold proximal hub ¿ it would only advance to the area of the balloon chamber where the guidewire lumen was stretched.The balloon chamber was inflated to nominal pressure of 8atm.The dilatation catheter was then inflated to rated burst pressure of 14atm.The dilatation catheter was able to hold nominal and rated burst pressures.A vacuum was pulled, and the balloon chamber would not fully deflate at either one minute or 90 seconds.The nanocross elite dilatation catheter exhibited a narrowing of the distal edge of the inflation lumen.This narrowing of the inflation pathway would have occurred during manufacturing.If information is provided in the future, a supplemental report will be issued.
|